Accuris MW9600 Manuel utilisateur

I
MW9600 Microplate Washer
User Manual
Version 1.1
Accuris Instruments is a division of
Benchmark Scientific.

II
Thank you for purchasing our MW9600 Microplate Washer. This user manual
describes how the instrument functions, and also includes an operation guide
and important safety information, please read it carefully before operating and
keep this manual for future reference.
Initial check
Upon receipt, please check the outer package for damage, and check all parts
inside the package to make sure all parts and components are included and in
proper condition. If anything is missing or damaged, please contact your
distributor or Benchmark Scientific immediately.
Packaging
Please keep the box and packaging material for your microplate washer. In the
event that service is required, the box will be needed for shipping the instrument
to our Service Department.

III
Safety warning and guidelines
1 Important guidelines
2 Safety
In order to avoid injury, please read this manual completely before operating
the instrument. Pay special attention to all safety tips.
This product is for indoor use only.
This instrument should only be operated by trained laboratory personnel.
Do not attempt to repair the instrument without receiving specific instructions
from Bechmark Scientific’s Service Personnel. Please contact your
distributor or Benchmark Scientific to arrange service or repair.
This instrument is intended for research only.
Warning: Risk of biological contamination. All samples used for testing,
quality control or calibration should be considered potentially infectious,
and any parts of the instrument that contact the samples should also be
considered at potentially infectious. Always wear gloves and appropriate
safety gear when operating this device.
Never place a hand on or near any moving parts when the instrument is
running.

IV
3 Maintenance
The plate holder and platform should be cleaned regularly with alcohol to remove any
dried solutions or debris.
The Rinse function should be used to wash the tubing and wash heads. If the wash
head’s metal tubed become clogged, they can be cleared using the included cleaning tool.
This instrument should be operated in a low humidity, low dust environment,
away from direct sunlight, heat and magnetic fields.
Always make sure to use the correct input voltage. If the power cable is
damaged, replace it with same type.
Turn off the power and unplug the instrument if not being used for a long period
of time. All bottles should and tubing should be rinsed and emptied. When
storing, cover the instrument to protect it from dust.
Shut off the power immediately if any of the following occur and then contact your
distributor of Benchmark Scientific’s Service Department:
★Fluid enters into the instrument
★The instrument is exposed to rain or water
★There are any unusual noises or smells coming from the instrument
★The instrument is dropped or the outer housing is broken
★There are any obvious malfunctions

V
CONTENTS
Chapter 1 Introduction...................................................................................................1
Chapter 2 Specifications...............................................................................................2
Chapter 3 Instrument Overview.................................................................................. 3
Chapter 4 Installation.....................................................................................................5
Chapter 5 Start Up and Software................................................................................ 8
Chapter 6 Preparing Wash and Waste Bottles......................................................17
Chapter 7 Creating a new wash protocol............................................................... 18
Chapter 8 Running a wash protocol........................................................................ 22
Chapter 9 Importing, Exporting and Deleting Programs....................................25
Chapter 10 Power Off...................................................................................................28
Chapter 11 Maintenance............................................................................................. 29
Chapter 12 Trouble Shooting.....................................................................................30
Appendix:MW9600 diagram.....................................................................................31

Microplate Washer User Manual Chapter 1 Introduction
─1─
Chapter 1 Introduction
The MW9600 Microplate Washer is an accessory to a Microplate Reader, and is designed
for accurate and efficient washing and preparation of ELISA plates prior to reading. The
MW9600 is compatible with standard 96-well, SBS-footprint microplates.
Features:
4.3 inch LCD display for easy programming and operating.
Reliable, high efficiency liquid pump
8 and 12-needle wash heads, suitable for micro plates with Flat, U, V, and C type well
bottoms.
Wash head positions are accurately controlled to +/- 0.1mm.
Multiple wash programs can be set up, stored and selected.
Program allows selecting specific rows for washing.
Pause function allows pausing the currently running program.
Wash bottles include clearly marked volume graduations to visually monitor liquid
levels.
Automatic monitoring and warnings when wash liquid levels are low or if waste liquid
level is high.
Efficient aspiration of liquid from plate wells, remaining liquid per well: ≤1uL.
Plate platform includes a beveled spill well. Any spilled or leaked fluid is extracted
automatically to avoid cross contamination.

Microplate Washer User Manual Chapter 2 Specifications
─2─
Chapter 2 Specifications
Operating conditions
Operating temperature range:4C45C
Operating relative humidity:≤70%
Power supply:AC 90-240V 50/60Hz
Parameters and features
Parameters Model
MW9600
Wash heads
8 channel or 12 channel
Wash modes
1 point,2 point,or 3 point aspiration
Compatible Micro
plates
96 wells, standard SBS dimensions
Compatible Well
bottoms
Flat, U, V, C bottom
Residual liquid
volume
≤1uL(each well)
Dispensing Volume
Range
50uL~2000uL
Wash cycles
1~99
Programs
100
Rinse function time
range
00:00~03:59 (mm: ss)
Soak and shaking
time range
00:00~99:00 (mm: ss)
Wash bottle level
warning
Yes, sensors included in wash and rinse bottles
Wash solution
channels
3 bottles
Waste bottle level
warning
Yes, sensor detects when bottle is full
Waste aspiration
Yes
Dimensions
(WXDXH)
345 x 460 x 200mm / 13.5 x 18 x 8.5 inches
Net weight
11.5kg / 25.3 Lbs

Microplate Washer User Manual Chapter 3 Instrument Overview
─3─
Chapter 3 Instrument Overview
Wash bottle A
Wash bottle B
Wash bottle R
Waste bottle W
Transparent Cover
Wash
LCD display
Display housing
Rear housing
Plate holder
Platform

Microplate Washer User Manual Chapter 3 Instrument Overview
─4─
Connect a flash drive to the USB port to transfer programs between
different MW9600 washers.
Wash head 2 wash heads available:1X8 channel and 1X12 channel:
1X12 channel wash head
USB port
1X8 channel wash head
USB port
SD card slot
Waste W exit
Waste W
negative pressure
Bottle R inlet
Bottle B inlet
Bottle A inlet
Bottle A fluid level warning
Bottle B fluid level warning
Bottle R fluid level warning
Bottle W fluid level warning
Power switch

Microplate Washer User Manual Chapter 4 Installation
─5─
Chapter 4 Installation
Initial Inspection
Thoroughly check the package and instrument upon receipt. Please contact your distributor or
Benchmark Scientific immediately if:
The package has been damaged or deformed during shipping.
The packaging is wet.
Any parts or accessories are missing (compare to the packing list). Make sure all ordered
items are included.
There is any damage to the Microplate Washer or accessories.
Installation steps
Open the package, remove the four bottles, wash heads, power supply and the instrument
and place in onto a sturdy, flat and level desk or lab bench.
Attention: Remove the shipping safety lock before operating.
Unscrew the allen screw on the wash head rack using the included allen
key.